Search

Cart

0 products 0,00 Kč Open cart >>

Products

Information

  Introduction     Products     Our dealers     Order     About us     Contact     Special  

Flash 06S DAD 600 Detector in housing

Information
click for detailclick for detail Catalog number: 2050S00000

Four wavelengths simultaneously,
scan of whole spectrum 200 - 600 nm

Description

Flash 06S DAD 600 is an UV diode array detector, which allows measuring absorbance of four wavelengths simultaneously in one cell just as measuring of whole spectrum (scan). This unit is used in liquid chromatography to verify analyzed samples by means of four wavelengths or in situations when some peaks absorb on different wavelengths. It is possible to use the detector in flash and preparative applications. By using analytical cell can be detector used also for analytical purpose. Noise level at 254 nm is ± 5 x 10-5 AU with a test cell.

The unit is designed as stand alone with 90–264 V power supply and RS232 controlling. It is derived from OEM build-in unit with powering 24-36 V DC.

The unit’s DAD (diode array detector) design offers many advantages:

  • wavelength setting from 200 up to 600 nm in increments of 1nm
  • absorbance measuring on four wavelengths simultaneously
  • scan of whole spectrum is possible in each time
  • lamp work hours are counted using the built-in counter
  • the cell is easy to replace from the side of the detector
  • powering by 90–264 V
  • unit is controlled by RS232 interface
Specification

Technical parameters:

Wavelength range 200 - 600 nm (256 elements on CCD)
Typical spectral half-width 8 nm
Accuracy of adjustment ± 1 nm
Reproducibility ± 0.5 nm
Light source Deuterium discharge lamp
Noise level at test cell (254 nm, TC 0.75 s) ± 5 x 10-5 AU
Drift at test cell (254 nm after 1 hr) 1 x 10-3 AU/hr
Materials in contact with mobile phase PTFE; fused silica, stainless steel
Time constant 0.5 s; 0.75 s; 1 s or 2 s
Output for integrator 1 V/AU (in digital form only)
Interface RS 232
Power supply 90–264 V
Power input 90 VA
Dimensions (W x H x D) 220 x 170 x 450 mm
Weight 7.9 kg

Preparative cell PLCC 15 (delivered with unit):

volume/optical path(adjustable) 45 μl / 0.3 mm (basic)
55 μl / 1.4 mm
70 μl / 2.4 mm
connecting tubing with OD = 1/8", thread 1/4"-28
maximal flow rate 500 ml/min
Count: pcs